User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Genisco GF-4150-2 Capacitor INSTRUMENT: Solar 7113-106R Capacitor DATE: 23-Nov-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 5 NUMBER OF LINES: 67 STANDARD: esi 273 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F 1.002 ASK+ X 1.003 HEAD * INITIAL CONDITIONS * 1.004 DISP The following Standard test equipment is required: 1.004 DISP esi 273 Capacitance Bridge 1.005 JMPT 1.008 Solar 7113-106R Capacitor 1.006 DISP Set the esi 273 front panel controls as follows: 1.006 DISP POWER ON/OFF ...................................... ON 1.006 DISP FREQUENCY kHz .................................. 0.120 1.006 DISP GEN LEVEL ................................... fully cw 1.006 DISP DISIPATION FACTOR dials .................... 0.001 x 0 1.006 DISP C RANGE .......................................... 1µF 1.006 DISP CAPACITANCE dials .............................. 9.999 1.006 DISP LOG/LIN switch ................................... LOG 1.007 JMP 1.009 1.008 DISP Set the esi 273 front panel controls as follows: 1.008 DISP POWER ON/OFF ...................................... ON 1.008 DISP FREQUENCY kHz .................................. 0.120 1.008 DISP GEN LEVEL ................................... fully cw 1.008 DISP DISIPATION FACTOR dials .................... 0.001 x 0 1.008 DISP C RANGE ......................................... 10µF 1.008 DISP CAPACITANCE dials .............................. 1.000 1.008 DISP LOG/LIN switch ................................... LOG 1.009 DISP Set the DETECTOR SENSITIVITY control so that line on 1.009 DISP knob lines up with dot at 9 o'clock. 1.010 PIC esi 1.011 DISP Adjust CAPACITANCE and DISSIPATION FACTOR dials 1.011 DISP alternately for a null on the meter. 1.012 DISP Set the DETECTOR SENSITIVITY control fully cw. 1.013 DISP Adjust CAPACITANCE and DISSIPATION FACTOR dials 1.013 DISP alternately for a null on the meter. 1.014 HEAD {* CAPACITANCE TEST *} 1.015 JMP 2.001 1.016 EVAL 2.001 STD esi 273 2.002 MEMI Use keyboard to enter 273 reading in µf: 2.003 ACC 10.000µf 0.1% 0.001U 2.004 MEME 2.005 MEMC µf 1.000U #! Test Tol 1, Sys Tol 0.011, TUR 90.9091 (>= 4.00). 3.001 HEAD {* DISSIPATION FACTOR TEST *} 3.002 JMP 4.001 3.003 EVAL 4.001 MEMI Use keyboard to enter 273 Dissipation Factor (red scale) 4.002 ACC 0.000 0.0011U +1% 4.003 MEME 4.004 MEMC D.F. +1.000U #! Test Tol 1, Sys Tol 0.0011, TUR 909.091 (>= 4.00). 5.001 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.